Different methods of correcting the readings of the DVGN-01 albedo neutron dosimeter placed behind the IBR-2M protection

Authors

  • Yu. V. Mokrov Author
  • S.V. Morozova Author
  • G.N. Timoshenko Author
  • V.A. Krylov Author
  • O.B. Badunov Author
  • G.A. Seilkhanova Author

Abstract

The paper presents the results of correction of readings of DVBN-01 albedo dosimeters behind IBR-2M protection, the Neutron Physics Laboratory of the Joint Institute for Nuclear Research (JINR), Dubna, byvarious methods. The neutron spectra were measured at two points behind the IBR-2M protection in the experimental halls, and measurements were carried out with the spherical albedo system at these points. The correction coefficients for DVGN-01 were calculated from the measured spectra based on the data of the spherical albedo system. A good agreement of the coefficients calculated by different methods is shown, which indicates the reliability of the obtained values for the correction coefficients. Based on the results of this work and the data obtained in the other investigations, the values of correction factors
recommended for use in individual radiation control (IRC) in the FLNP were obtained.

Recommended Citation

Mokrov, Yu. V.; Morozova, S. V.; Timoshenko, G. N.; Krylov, V. A.; Badunov, O. B.; and Seilkhanova, G. A. (2018) "Different methods of correcting the readings of the DVGN-01 albedo neutron dosimeter placed behind the IBR-2M protection," Eurasian Journal of Physics and Functional Materials: Vol. 2: No. 4, Article 2.
DOI: https://doi.org/10.29317/ejpfm.2018020402

References

Downloads

Published

2026-03-03

How to Cite

(1)
Mokrov, Y.; Morozova, S.; Timoshenko, G.; Krylov, V.; Badunov, O.; Seilkhanova, G. Different Methods of Correcting the Readings of the DVGN-01 Albedo Neutron Dosimeter Placed Behind the IBR-2M Protection. Eur. J. Phys. Funct. Mater. 2026, 2 (4).