Different methods of correcting the readings of the DVGN-01 albedo neutron dosimeter placed behind the IBR-2M protection
Abstract
The paper presents the results of correction of readings of DVBN-01 albedo dosimeters behind IBR-2M protection, the Neutron Physics Laboratory of the Joint Institute for Nuclear Research (JINR), Dubna, byvarious methods. The neutron spectra were measured at two points behind the IBR-2M protection in the experimental halls, and measurements were carried out with the spherical albedo system at these points. The correction coefficients for DVGN-01 were calculated from the measured spectra based on the data of the spherical albedo system. A good agreement of the coefficients calculated by different methods is shown, which indicates the reliability of the obtained values for the correction coefficients. Based on the results of this work and the data obtained in the other investigations, the values of correction factors
recommended for use in individual radiation control (IRC) in the FLNP were obtained.
Recommended Citation
Mokrov, Yu. V.; Morozova, S. V.; Timoshenko, G. N.; Krylov, V. A.; Badunov, O. B.; and Seilkhanova, G. A. (2018) "Different methods of correcting the readings of the DVGN-01 albedo neutron dosimeter placed behind the IBR-2M protection," Eurasian Journal of Physics and Functional Materials: Vol. 2: No. 4, Article 2.
DOI: https://doi.org/10.29317/ejpfm.2018020402
References
Downloads
Published
Issue
Section
License
The Eurasian Journal of Physics and Functional Materials is an open-access journal. Articles published under the Creative Commons Attribution 4.0 International License (CC BY 4.0) may be shared, redistributed, and adapted, including for commercial purposes, provided that the terms of the license are followed, including appropriate attribution to the original authors and source.
License: https://creativecommons.org/licenses/by/4.0/
Copyright ownership and publication rights are governed by the terms applicable to each article and the relevant author agreement.


